Field-Amplified Sample Injection-Micellar Electrokinetic Chromatography for the Determination of Benzophenones in Food Simulants

Fecha de publicación

2015-09-18T11:22:21Z

2015-09-18T11:22:21Z

2015-07-20

2015-09-18T11:22:21Z

Resumen

A field-amplified sample injection-micellar electrokinetic chromatography (FASI-MEKC) method for the determination of 14 benzophenones (BPs) in a food simulant used in migration studies of food packaging materials was developed, allowing almost baseline separation in less than 21 min. The use of a 10 mM sodium dodecyl sulfate (SDS) solution as sample matrix was mandatory to achieve FASI enhancement of the analyzed BPs. A 21- to 784-fold sensitivity enhancement was achieved with FASI-MEKC, obtaining limits of detection down to 5.1<br>68.4 µg/L, with acceptable run-to-run precisions (RSD values lower than 22.3%) and accuracy (relative errors lower than 21.0%). Method performance was evaluated by quantifying BPs in the food simulant spiked at 500 µg/L (bellow the established specific migration limit for BP (600 µg/L) by EU legislation). For a 95% confidence level, no statistical differences were observed between found and spiked concentrations (probability at the confidence level, p value, of 0.55), showing that the proposed FASI-MEKC method is suitable for the analysis of BPs in food packaging migration studies at the levels established by EU legislation.

Tipo de documento

Artículo


Versión publicada

Lengua

Inglés

Publicado por

MDPI

Documentos relacionados

Reproducció del document publicat a: http://dx.doi.org/10.3390/chromatography2030452

Chromatography, 2015, vol. 2(3), p. 452-471

http://dx.doi.org/10.3390/chromatography2030452

Citación recomendada

Esta citación se ha generado automáticamente.

Derechos

cc-by (c) Félez, Cristina et al., 2015

http://creativecommons.org/licenses/by/3.0/es

Este ítem aparece en la(s) siguiente(s) colección(ones)