An FIA-MS Method for Rapid Coffee Adulteration Detection: A Comparative Study with a Non-Targeted LC-MS Approach

Publication date

2025-09-19T16:35:15Z

2025-09-19T16:35:15Z

2025-08

2025-09-19T16:35:15Z

Abstract

Coffee adulteration is a growing concern in the food industry due to economic and quality implications. This study evaluates a rapid, non-targeted fingerprinting method based on flow injection analysis–mass spectrometry (FIA-MS) for detecting common coffee adulterants. A total of 119 samples were analyzed, including 43 coffee samples and 76 samples of common coffee adulterants (16 chicory, 10 barley, and 50 flour samples). FIA-MS combined with chemometric analysis allowed for the classification of pure and adulterated coffee samples with over 95% accuracy. Compared to LC-MS, the FIA-MS method showed a similar performance while offering significantly faster analysis and lower solvent consumption, making it a practical and sustainable option for high-throughput screening. For PLS regression studies, calibration and prediction errors were consistently below 0.91% and 11.7%, respectively. Furthermore, the methodology was compared with a non-targeted LC-MS approach, showing an excellent performance.

Document Type

Article


Published version

Language

English

Publisher

MDPI

Related items

Reproducció del document publicat a: https://doi.org/https://doi.org/10.3390/foods14172931

Foods, 2025, vol. 14, num.2391

https://doi.org/https://doi.org/10.3390/foods14172931

Recommended citation

This citation was generated automatically.

Rights

cc-by (c) Núñez et al., 2025

http://creativecommons.org/licenses/by/4.0/