Image Quality of Data Products of the High Resolution Telescope of the Polarimetric and Helioseismic Imager

Resumen

The High Resolution Telescope (HRT) of the Polarimetric and Helioseismic Imager (SO/PHI) on-board the Solar Orbiter mission (SO) provides near diffraction limited observations of the solar surface. The HRT Refocus Mechanism (HRM) allows for acquiring calibration data in flight which are used in post processing on ground to estimate the image quality of SO/PHI-HRT data products and its dependence on the SO-Sun distance. Our aim is to characterise the wavefront aberrations in the optical path of SO/PHI-HRT and consequently the image quality in the focal plane of the telescope. We use calibration data taken during the Near Earth Commissionning Phase (NECP) and the second Remote Sensing Check-out Window (RSCW2) of Solar Orbiter’s Cruise Phase (CP). In particular, we apply a Phase Diversity (PD) analysis to estimate the low-order wavefront aberrations. The restoration with the retrieved Point Spread Function (PSF) from the PD analysis increases the RMS contrast of the solar granulation in the visible continuum from 4 % to 10−11%.

Tipo de documento

Artículo


Versión publicada

Lengua

Inglés

Publicado por

Society of Photo-Optical Instrumentation Engineers (SPIE)

Documentos relacionados

Reproducció del document publicat a: https://doi.org/10.1117/12.2628942

Proceedings of SPIE, 2022, vol. 12180

https://doi.org/10.1117/12.2628942

Citación recomendada

Esta citación se ha generado automáticamente.

Derechos

(c) Society of Photo-Optical Instrumentation Engineers (SPIE), 2022

Este ítem aparece en la(s) siguiente(s) colección(ones)