dc.description.abstract
GC-IMS is a very good complementary technique to traditional GC-MS, that presents some advantages, but also, some disadvantages such as misalignments produced by many parameters affecting the equipment stability. The reproducibility of the measures has been studied in two different measurement campaigns with a set of automatized quality control parameters. Figures of merit from one region of interest present in the samples show that the saturation and asymmetry do not change between measurement campaigns, but the volume and area of the total ion spectra change. A correction of these changes between batches should be developed.
dc.format
application/pdf
dc.relation
Versió postprint del document publicat a: https://doi.org/10.1109/ISOEN54820.2022.9789683
dc.relation
Comunicació a: 2022 IEEE International Symposium on Olfaction and Electronic Nose (ISOEN), International Society for Olfaction and Chemical Sensing (ISOCS), Aveiro, Portugal, 29 de maig - 1 de juny 2022
dc.relation
https://doi.org/10.1109/ISOEN54820.2022.9789683
dc.rights
info:eu-repo/semantics/openAccess
dc.source
Comunicacions a congressos (Enginyeria Electrònica i Biomèdica)
dc.subject
Mesurament de superfícies
dc.subject
Control de qualitat
dc.subject
Area measurement
dc.subject
Quality control
dc.title
Study of quality controls for stability check of the ROIs of a ketones mixture in different GC-IMS measurement campaigns
dc.type
info:eu-repo/semantics/conferenceObject
dc.type
info:eu-repo/semantics/acceptedVersion