We present a simple an inexpensive method to implement a Griess-Saltzman-type reaction that combines the advantages of the liquid phase method (high specificity, fast response time) with the benefits of a solid implementation (easy to handle). We demonstrate that the measurements can be carried out using conventional RGB sensors; circumventing all the limitations around the measurement of the samples with spectrometers. We also present a method to optimize the measurement protocol and target a specific range of NO2 concentrations. We demonstrate that it is possible to measure the concentration of NO2 from 50 ppb to 300 ppm with high specificity and without modifying the sensing elements.
Article
Versió publicada
Anglès
Colorimetria; Xarxes de sensors; Colorimetry; Sensor networks
MDPI
Reproducció del document publicat a: https://doi.org/10.3390/proceedings1040471
MDPI Proceedings, 2017, vol. 1, num. 4, p. 471
https://doi.org/10.3390/proceedings1040471
cc-by (c) Fernández, Luisa et al., 2017
https://creativecommons.org/licenses/by/4.0/