Simultaneous voltammetric determination of heavy metals by use of crown ether-modified electrodes and chemometrics

Data de publicació

2019-09-20T08:13:15Z

2019-09-20T08:13:15Z

2016-10-12

2019-09-20T08:13:16Z

Resum

A three‐sensor array consisting of a graphite‐epoxy composite electrode (GEC), 4‐carboxybenzo‐18‐crown‐6‐GEC and 4‐carboxybenzo‐15‐crown‐5‐GEC was employed for the simultaneous determination of Cd(II), Pb(II) and Hg(II) by differential pulse anodic stripping voltammetry (DPASV). Sensors were firstly studied for the determination of Hg(II); secondly, peak current responses confirmed that all sensors showed differentiated response for the three considered metals. A response model was developed to resolve mixtures of Cd(II), Pb(II) and Hg(II) at the µg L−1 level; Discrete Wavelet Transform was selected as preprocessing tool and artificial neural network used for the modelling of the obtained responses.

Tipus de document

Article


Versió acceptada

Llengua

Anglès

Publicat per

Wiley-VCH

Documents relacionats

Versió postprint del document publicat a: https://doi.org/10.1002/elan.201500512

Electroanalysis, 2016, vol. 28, p. 663-670

https://doi.org/10.1002/elan.201500512

Citació recomanada

Aquesta citació s'ha generat automàticament.

Drets

(c) Wiley-VCH, 2016

Aquest element apareix en la col·lecció o col·leccions següent(s)