We present a simple and inexpensive method to implement a Griess-Saltzman-type reaction that combines the advantages of the liquid phase method (high specificity and fast response time) with the benefits of a solid implementation (easy to handle). We demonstrate that the measurements can be carried out using conventional RGB sensors; circumventing all the limitations around the measurement of the samples with spectrometers. We also present a method to optimize the measurement protocol and target a specific range of NO2 concentrations. We demonstrate that it is possible to measure the concentration of NO2 from 50 ppb to 300 ppm with high specificity and without modifying the Griess-Saltzman reagent.
Article
Accepted version
English
Colorimetria; Medi ambient; Colorimetry; Natural environment
American Chemical Society
Versió postprint del document publicat a: https://doi.org/10.1021/acssensors.7b00463
ACS Sensors, 2017, vol. 2, num. 11, p. 1612-1618
https://doi.org/10.1021/acssensors.7b00463
info:eu-repo/grantAgreement/EC/FP7/336917/EU//BETTERSENSE
info:eu-repo/grantAgreement/EC/H2020/640525/EU//GasApp
(c) American Chemical Society , 2017