A Point-of-Care Device for Molecular Diagnosis Based on CMOS SPAD Detectors with Integrated Microfluidics

dc.contributor.author
Canals Gil, Joan
dc.contributor.author
Franch Masdeu, Nil
dc.contributor.author
Alonso Casanovas, Oscar
dc.contributor.author
Vilà i Arbonès, Anna Maria
dc.contributor.author
Diéguez Barrientos, Àngel
dc.date.issued
2021-04-08T10:02:43Z
dc.date.issued
2021-04-08T10:02:43Z
dc.date.issued
2019-01-22
dc.date.issued
2021-04-08T10:02:43Z
dc.identifier
1424-8220
dc.identifier
https://hdl.handle.net/2445/176049
dc.identifier
689473
dc.identifier
30678225
dc.description.abstract
We describe the integration of techniques and technologies to develop a Point-of-Care for molecular diagnosis PoC-MD, based on a fluorescence lifetime measurement. Our PoC-MD is a low-cost, simple, fast, and easy-to-use general-purpose platform, aimed at carrying out fast diagnostics test through label detection of a variety of biomarkers. It is based on a 1-D array of 10 ultra-sensitive Single-Photon Avalanche Diode (SPAD) detectors made in a 0.18 μm High-Voltage Complementary Metal Oxide Semiconductor (HV-CMOS) technology. A custom microfluidic polydimethylsiloxane cartridge to insert the sample is straightforwardly positioned on top of the SPAD array without any alignment procedure with the SPAD array. Moreover, the proximity between the sample and the gate-operated SPAD sensor makes unnecessary any lens or optical filters to detect the fluorescence for long lifetime fluorescent dyes, such as quantum dots. Additionally, the use of a low-cost laser diode as pulsed excitation source and a Field-Programmable Gate Array (FPGA) to implement the control and processing electronics, makes the device flexible and easy to adapt to the target label molecule by only changing the laser diode. Using this device, reliable and sensitive real-time proof-of-concept fluorescence lifetime measurement of quantum dot QdotTM 605 streptavidin conjugate is demonstrated.
dc.format
18 p.
dc.format
application/pdf
dc.language
eng
dc.publisher
MDPI
dc.relation
Reproducció del document publicat a: https://doi.org/10.3390/s19030445
dc.relation
Sensors, 2019, vol. 19, num. 3, p. 445
dc.relation
https://doi.org/10.3390/s19030445
dc.relation
info:eu-repo/grantAgreement/EC/H2020/737089/EU//ChipScope
dc.rights
cc-by (c) Canals Gil, Joan et al., 2019
dc.rights
http://creativecommons.org/licenses/by/3.0/es
dc.rights
info:eu-repo/semantics/openAccess
dc.source
Articles publicats en revistes (Enginyeria Electrònica i Biomèdica)
dc.subject
Diagnòstic molecular
dc.subject
Fluorescència
dc.subject
Díodes
dc.subject
Molecular diagnosis
dc.subject
Fluorescence
dc.subject
Diodes
dc.title
A Point-of-Care Device for Molecular Diagnosis Based on CMOS SPAD Detectors with Integrated Microfluidics
dc.type
info:eu-repo/semantics/article
dc.type
info:eu-repo/semantics/publishedVersion


Files in this item

FilesSizeFormatView

There are no files associated with this item.

This item appears in the following Collection(s)