Evaluation of an Artificial Intelligence-Based Tool and a Universal Low-Cost Robotized Microscope for the Automated Diagnosis of Malaria

Otros/as autores/as

Institut Català de la Salut

[Rubio Maturana C, Mediavilla A, Martínez-Vallejo P, Goterris L] Servei de Microbiologia, Vall d’Hebron Hospital Universitari, Barcelona, Spain. Vall d’Hebron Institut de Recerca (VHIR), Barcelona, Spain. Department of Microbiology and Genetics, Universitat Autònoma de Barcelona (UAB), Bellaterra, Spain. [Dantas de Oliveira A] Computational Biology and Complex Systems Group, Physics Department, Universitat Politècnica de Catalunya (UPC), Castelldefels, Spain. [Zarzuela F, Muixí M, Joseph-Munné J] Servei de Microbiologia, Vall d’Hebron Hospital Universitari, Barcelona, Spain. Vall d’Hebron Institut de Recerca (VHIR), Barcelona, Spain. [Silgado A, Sulleiro E] Servei de Microbiologia, Vall d’Hebron Hospital Universitari, Barcelona, Spain. Vall d’Hebron Institut de Recerca (VHIR), Barcelona, Spain. Department of Microbiology and Genetics, Universitat Autònoma de Barcelona (UAB), Bellaterra, Spain. Centro de Investigación Biomédica en Red Enfermedades

Vall d'Hebron Barcelona Hospital Campus

Fecha de publicación

2025-03-14T13:29:33Z

2025-03-14T13:29:33Z

2024

2025



Resumen

Artificial intelligence; Automated diagnosis; Malaria


Inteligencia artificial; Diagnóstico automatizado; Malaria


Intel·ligència artificial; Diagnòstic automatitzat; Malària


The gold standard diagnosis for malaria is the microscopic visualization of blood smears to identify Plasmodium parasites, although it is an expert-dependent technique and could trigger diagnostic errors. Artificial intelligence (AI) tools based on digital image analysis were postulated as a suitable supportive alternative for automated malaria diagnosis. A diagnostic evaluation of the iMAGING AI-based system was conducted in the reference laboratory of the International Health Unit Drassanes-Vall d’Hebron in Barcelona, Spain. iMAGING is an automated device for the diagnosis of malaria by using artificial intelligence image analysis tools and a robotized microscope. A total of 54 Giemsa-stained thick blood smear samples from travelers and migrants coming from endemic areas were employed and analyzed to determine the presence/absence of Plasmodium parasites. AI diagnostic results were compared with expert light microscopy gold standard method results. The AI system shows 81.25% sensitivity and 92.11% specificity when compared with the conventional light microscopy gold standard method. Overall, 48/54 (88.89%) samples were correctly identified [13/16 (81.25%) as positives and 35/38 (92.11%) as negatives]. The mean time of the AI system to determine a positive malaria diagnosis was 3 min and 48 s, with an average of 7.38 FoV analyzed per sample. Statistical analyses showed the Kappa Index = 0.721, demonstrating a satisfactory correlation between the gold standard diagnostic method and iMAGING results. The AI system demonstrated reliable results for malaria diagnosis in a reference laboratory in Barcelona. Validation in malaria-endemic regions will be the next step to evaluate its potential in resource-poor settings.


This research was co-funded by the Microbiology Department of Vall d’Hebron University Hospital and the Cooperation Centre of the Universitat Politècnica de Catalunya (CCD-UPC) and Probitas Foundation.

Tipo de documento

Artículo


Versión publicada

Lengua

Inglés

Publicado por

MDPI

Documentos relacionados

International Journal of Environmental Research and Public Health;22(1)

https://doi.org/10.3390/ijerph22010047

Citación recomendada

Esta citación se ha generado automáticamente.

Derechos

Attribution 4.0 International

http://creativecommons.org/licenses/by/4.0/

Este ítem aparece en la(s) siguiente(s) colección(ones)