<?xml version="1.0" encoding="UTF-8"?><?xml-stylesheet type="text/xsl" href="static/style.xsl"?><OAI-PMH xmlns="http://www.openarchives.org/OAI/2.0/" xmlns:xsi="http://www.w3.org/2001/XMLSchema-instance" xsi:schemaLocation="http://www.openarchives.org/OAI/2.0/ http://www.openarchives.org/OAI/2.0/OAI-PMH.xsd"><responseDate>2026-04-17T13:53:49Z</responseDate><request verb="GetRecord" identifier="oai:www.recercat.cat:2445/8686" metadataPrefix="oai_dc">https://recercat.cat/oai/request</request><GetRecord><record><header><identifier>oai:recercat.cat:2445/8686</identifier><datestamp>2025-11-20T13:50:42Z</datestamp><setSpec>com_2072_1057</setSpec><setSpec>col_2072_478799</setSpec><setSpec>col_2072_478917</setSpec></header><metadata><oai_dc:dc xmlns:oai_dc="http://www.openarchives.org/OAI/2.0/oai_dc/" xmlns:dc="http://purl.org/dc/elements/1.1/" xmlns:xsi="http://www.w3.org/2001/XMLSchema-instance" xmlns:doc="http://www.lyncode.com/xoai" xsi:schemaLocation="http://www.openarchives.org/OAI/2.0/oai_dc/ http://www.openarchives.org/OAI/2.0/oai_dc.xsd">
   <dc:title>A new estimator to minimize the error due to breathing in the measurement of respiratory impedance</dc:title>
   <dc:creator>Navajas Navarro, Daniel</dc:creator>
   <dc:creator>Farré Ventura, Ramon</dc:creator>
   <dc:creator>Rotger i Estapé, Maria del Mar</dc:creator>
   <dc:creator>Peslin, René</dc:creator>
   <dc:subject>Errors científics</dc:subject>
   <dc:subject>Mesurament</dc:subject>
   <dc:subject>Biomedical measurement</dc:subject>
   <dc:subject>Measurement errors</dc:subject>
   <dc:subject>Pneumodynamics</dc:subject>
   <dc:description>This paper presents a new respiratory impedance estimator to minimize the error due to breathing. Its practical reliability was evaluated in a simulation using realistic signals. These signals were generated by superposing pressure and flow records obtained in two conditions: 1) when applying forced oscillation to a resistance- inertance- elastance (RIE) mechanical model; 2) when healthy subjects breathed through the unexcited forced oscillation generator. Impedances computed (4-32 Hz) from the simulated signals with the new estimator resulted in a mean value which was scarcely biased by the added breathing (errors less than 1 percent in the mean R, I , and E ) and had a small variability (coefficients of variation of R, I, and E of 1.3, 3.5, and 9.6 percent, respectively). Our results suggest that the proposed estimator reduces the error in measurement of respiratory impedance without appreciable extracomputational cost.</dc:description>
   <dc:date>2009-06-17T08:40:18Z</dc:date>
   <dc:date>2009-06-17T08:40:18Z</dc:date>
   <dc:date>1988</dc:date>
   <dc:type>info:eu-repo/semantics/article</dc:type>
   <dc:type>info:eu-repo/semantics/publishedVersion</dc:type>
   <dc:identifier>0018-9294</dc:identifier>
   <dc:identifier>https://hdl.handle.net/2445/8686</dc:identifier>
   <dc:identifier>35251</dc:identifier>
   <dc:language>eng</dc:language>
   <dc:relation>Reproducció del document publicat a http://dx.doi.org/10.1109/10.8684</dc:relation>
   <dc:relation>IEEE Transactions on Biomedical Engineering, 1988, vol. 35, núm. 12, p. 1001-1005.</dc:relation>
   <dc:relation>http://dx.doi.org/10.1109/10.8684</dc:relation>
   <dc:rights>(c) IEEE, 1988</dc:rights>
   <dc:rights>info:eu-repo/semantics/openAccess</dc:rights>
   <dc:format>5 p.</dc:format>
   <dc:format>application/pdf</dc:format>
   <dc:format>application/pdf</dc:format>
   <dc:publisher>IEEE</dc:publisher>
   <dc:source>Articles publicats en revistes (Ciències Fisiològiques)</dc:source>
</oai_dc:dc></metadata></record></GetRecord></OAI-PMH>